CVE-2014-5431
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
Published:Mar 26, 2019
Last Modified:Nov 21, 2024
EPS:Mar 26, 2019
EPSS Score:0.00053
CVSS Score:6.8
Affected Products
Vendor
Product
Action
Vendor
Baxter
Product
Sigma Spectrum Infusion System
Baxter
Sigma Spectrum Infusion System
Vendor
Baxter
Product
Sigma Spectrum Infusion System Firmware
Baxter
Sigma Spectrum Infusion System Firmware
Vendor
Baxter
Product
Wireless Battery Module
Baxter
Wireless Battery Module
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Related CVEs
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High