CVE-2015-3954
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published:Mar 25, 2019
Last Modified:Nov 21, 2024
EPS:Mar 25, 2019
EPSS Score:0.00537
CVSS Score:9.8
Affected Products
Vendor
Product
Action
Vendor
Pifzer
Product
Plum A\+3 Infusion System
Pifzer
Plum A\+3 Infusion System
Vendor
Pifzer
Product
Plum A\+3 Infusion System Firmware
Pifzer
Plum A\+3 Infusion System Firmware
Vendor
Pifzer
Product
Plum A\+ Infusion System
Pifzer
Plum A\+ Infusion System
Vendor
Pifzer
Product
Plum A\+ Infusion System Firmware
Pifzer
Plum A\+ Infusion System Firmware
Vendor
Pifzer
Product
Symbiq Infusion System
Pifzer
Symbiq Infusion System
Vendor
Pifzer
Product
Symbiq Infusion System Firmware
Pifzer
Symbiq Infusion System Firmware
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Related CVEs
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High