CVE Feed

    Dashboard / CVE / CVE-2017-11578

    CVE-2017-11578

    It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

    Published:Jul 2, 2019
    Last Modified:Nov 21, 2024
    EPS:Jul 2, 2019
    EPSS Score:0.01409
    CVSS Score:5.9

    Affected Products

    Vendor
    Blipcare
    Product
    Wi-fi Blood Pressure Monitor
    Vendor
    Blipcare
    Product
    Wi-fi Blood Pressure Monitor Firmware

    Common Attack Pattern Enumeration and Classification (CAPEC)

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High