    Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.

    Published:Aug 10, 2018
    Last Modified:May 19, 2026
    EPS:Aug 10, 2018
    EPSS Score:0.00036
    CVSS Score:4.4

    Affected Products

    Vendor
    Medtronic
    Product
    Mycarelink 24950 Patient Monitor
    Vendor
    Medtronic
    Product
    Mycarelink 24950 Patient Monitor Firmware
    Vendor
    Medtronic
    Product
    Mycarelink 24952 Patient Monitor
    Vendor
    Medtronic
    Product
    Mycarelink 24952 Patient Monitor Firmware

    Exploits

    No exploit reference

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High