CVE-2018-18563
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.
Published:Nov 20, 2018
Last Modified:Nov 21, 2024
EPS:Nov 20, 2018
EPSS Score:0.0025
CVSS Score:9.6
Affected Products
Vendor
Product
Action
Vendor
Roche
Product
Accu-chek Inform Ii
Roche
Accu-chek Inform Ii
Vendor
Roche
Product
Accu-chek Inform Ii Firmware
Roche
Accu-chek Inform Ii Firmware
Vendor
Roche
Product
Coaguchek Pro Ii
Roche
Coaguchek Pro Ii
Vendor
Roche
Product
Coaguchek Pro Ii Firmware
Roche
Coaguchek Pro Ii Firmware
Vendor
Roche
Product
Coaguchek Xs Plus
Roche
Coaguchek Xs Plus
Vendor
Roche
Product
Coaguchek Xs Plus Firmware
Roche
Coaguchek Xs Plus Firmware
Vendor
Roche
Product
Coaguchek Xs Pro
Roche
Coaguchek Xs Pro
Vendor
Roche
Product
Coaguchek Xs Pro Firmware
Roche
Coaguchek Xs Pro Firmware
Vendor
Roche
Product
Cobas H 232
Roche
Cobas H 232
Vendor
Roche
Product
Cobas H 232 Firmware
Roche
Cobas H 232 Firmware
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Related CVEs
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High