CVE Feed

    Dashboard / CVE / CVE-2019-25723

    CVE-2019-25723

    Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

    Published:Jun 2, 2026
    Last Modified:Jun 3, 2026
    EPS:Jun 2, 2026
    EPSS Score:0.00043
    CVSS Score:4

    Affected Products

    Vendor
    Draeger
    Product
    Perseus A500

    Exploits

    No exploit reference

    Common Attack Pattern Enumeration and Classification (CAPEC)

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High