CVE Feed

    Dashboard / CVE / CVE-2020-12037

    CVE-2020-12037

    Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

    Published:Jun 29, 2020
    Last Modified:Nov 21, 2024
    EPS:Jun 29, 2020
    EPSS Score:0.0007
    CVSS Score:7.5

    Affected Products

    Vendor
    Baxter
    Product
    Prismaflex
    Vendor
    Baxter
    Product
    Prismaflex Firmware
    Vendor
    Baxter
    Product
    Prismax
    Vendor
    Baxter
    Product
    Prismax Firmware

    Exploits

    No exploit reference

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High