CVE-2020-12048
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Published:Jun 29, 2020
Last Modified:Nov 21, 2024
EPS:Jun 29, 2020
EPSS Score:0.00061
CVSS Score:7.5
Affected Products
Vendor
Product
Action
Vendor
Baxter
Product
Phoenix X36
Baxter
Phoenix X36
Vendor
Baxter
Product
Phoenix X36 Firmware
Baxter
Phoenix X36 Firmware
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High