CVE-2021-33886
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Published:Aug 25, 2021
Last Modified:Nov 21, 2024
EPS:Aug 25, 2021
EPSS Score:0.00988
CVSS Score:8.1
Affected Products
Vendor
Product
Action
Vendor
Bbraun
Product
Infusomat Large Volume Pump 871305u
Bbraun
Infusomat Large Volume Pump 871305u
Vendor
Bbraun
Product
Spacecom2
Bbraun
Spacecom2
Vendor
Bbraun
Product
Spacestation 8713142u
Bbraun
Spacestation 8713142u
Exploits
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Related CVEs
References
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High