    CVE-2023-0888

    An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

    Published:Mar 13, 2023
    Last Modified:Feb 27, 2025
    EPS:Mar 13, 2023
    EPSS Score:0.00086
    CVSS Score:4.9

    Affected Products

    Vendor
    Bbraun
    Product
    Battery-pack Sp With Wifi
    Vendor
    Bbraun
    Product
    Battery-pack Sp With Wifi Firmware

    Exploits

    No exploit reference

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High