CVE Feed

    Dashboard / CVE / CVE-2024-9832

    CVE-2024-9832

    There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

    Published:Nov 14, 2024
    Last Modified:Apr 15, 2026
    EPS:Nov 14, 2024
    EPSS Score:0.00089
    CVSS Score:9.3

    Affected Products

    Vendor
    Baxter
    Product
    Life2000 Ventilator Firmware

    Exploits

    No exploit reference

    Related CVEs

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High