CVE-2024-9832
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Published:Nov 14, 2024
Last Modified:Apr 15, 2026
EPS:Nov 14, 2024
EPSS Score:0.00089
CVSS Score:9.3
Affected Products
Vendor
Product
Action
Vendor
Baxter
Product
Life2000 Ventilator Firmware
Baxter
Life2000 Ventilator Firmware
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
Related CVEs
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High