    A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

    Published:Dec 17, 2025
    Last Modified:Apr 15, 2026
    EPS:Dec 17, 2025
    EPSS Score:0.00286
    CVSS Score:7.2

    Affected Products

    Vendor
    Radiometer
    Product
    Abl800 Basic Analyzer
    Vendor
    Radiometer
    Product
    Abl800 Flex Analyzer
    Vendor
    Radiometer
    Product
    Abl90 Flex Analyzer
    Vendor
    Radiometer
    Product
    Abl90 Flex Plus Analyzer
    Vendor
    Radiometer
    Product
    Aqt90 Flex Analyzer

    Exploits

    No exploit reference

    Common Weakness Enumeration

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High