    CVE-2025-54304

    An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

    Published:Dec 4, 2025
    Last Modified:Dec 16, 2025
    EPS:Dec 4, 2025
    EPSS Score:0.0008
    CVSS Score:9.8

    Affected Products

    Vendor
    Thermofisher
    Product
    Ion Torrent Onetouch 2
    Vendor
    Thermofisher
    Product
    Ion Torrent Onetouch 2 Firmware

    Exploits

    No exploit reference

    Common Attack Pattern Enumeration and Classification (CAPEC)

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High