    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

    Published:May 26, 2026
    Last Modified:Jun 4, 2026
    EPS:May 26, 2026
    EPSS Score:0.00134
    CVSS Score:9.8

    Affected Products

    Vendor
    Eppendorf
    Product
    Bioflo 320

    Exploits

    No exploit reference

    Common Weakness Enumeration

    Common Attack Pattern Enumeration and Classification (CAPEC)

    No CAPEC recorded yet

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High