Smiths-medical

    Dashboard / Vendors

    Products: 2
    Vulnerabilities: 10
    Known Exploited: 0
    1
    Critical Level Threats
    5
    High Level Threats
    3
    Medium Level Threats
    1
    Low Level Threats

    Vulnerabilities

    100806040200
    JanFebMarAprMayJunJulAugSepOctNovDec
    Critical Level Threats
    High Level Threats
    Medium Level Threats
    Low Level Threats

    Products Security index

    Actions
    Items Per Page

    Vulnerabilities

    CVE-2017-12725

    A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

    Last Modified: Nov 21, 2024
    Published: Feb 15, 2018

    CVE-2017-12718

    A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

    Last Modified: Nov 21, 2024
    Published: Feb 15, 2018

    CVE-2017-12720

    An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.

    Last Modified: Nov 21, 2024
    Published: Feb 15, 2018

    CVE-2017-12721

    An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

    Last Modified: Nov 21, 2024
    Published: Feb 15, 2018

    CVE-2017-12724

    A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.

    Last Modified: Nov 21, 2024
    Published: Feb 15, 2018