Swisslog-healthcare

    Dashboard / Vendors

    Products: 2
    Vulnerabilities: 8
    Known Exploited: 0
    7
    Critical Level Threats
    1
    High Level Threats
    0
    Medium Level Threats
    0
    Low Level Threats

    Vulnerabilities

    100806040200
    JanFebMarAprMayJunJulAugSepOctNovDec
    Critical Level Threats
    High Level Threats
    Medium Level Threats
    Low Level Threats

    Products Security index

    Actions
    Items Per Page

    Vulnerabilities

    CVE-2021-37162

    A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.

    Last Modified: Nov 21, 2024
    Published: Aug 02, 2021

    CVE-2021-37164

    A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.

    Last Modified: Nov 21, 2024
    Published: Aug 02, 2021

    CVE-2021-37163

    An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded.

    Last Modified: Nov 21, 2024
    Published: Aug 02, 2021

    CVE-2021-37167

    An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the device, which provides permissions for all of the functionality of the device.

    Last Modified: Nov 21, 2024
    Published: Aug 02, 2021

    CVE-2021-37166

    A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.

    Last Modified: Nov 21, 2024
    Published: Aug 02, 2021
    Items Per Page