CVE-2017-12712

    Dashboard / Vulnerabilities / CVE-2017-12712

    CVE-2017-12712

    Published: 25 Apr 2018Last Modified: 10 Mar 2026

    Summary:

    Details: The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

    Affected packages

    Package

    Name:

    Purl:

    Affected ranges

    Type: N/A

    Events:

    Introduced- None
    Fixed -None

    Affected versions

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High