CVE-2017-12714
Dashboard / Vulnerabilities / CVE-2017-12714
CVE-2017-12714
Published: 25 Apr 2018Last Modified: 10 Mar 2026
Summary:
Details: Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References: , http://www.securityfocus.com/bid/100523, https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
Affected packages
Package
Name:
Purl:
Affected ranges
Type: N/A
Events:
Introduced- None
Fixed -None
Affected versions
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High