CVE-2017-12716
Dashboard / Vulnerabilities / CVE-2017-12716
CVE-2017-12716
Published: 25 Apr 2018Last Modified: 10 Mar 2026
Summary:
Details: Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References: , http://www.securityfocus.com/bid/100523, https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
Affected packages
Package
Name:
Purl:
Affected ranges
Type: N/A
Events:
Introduced- None
Fixed -None
Affected versions
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High