CVE-2021-33885
Dashboard / Vulnerabilities / CVE-2021-33885
CVE-2021-33885
Published: 25 Aug 2021Last Modified: 10 Mar 2026
Summary:
Details: An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
References: https://www.bbraunusa.com/en.htm, , https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Affected packages
Package
Name:
Purl:
Affected ranges
Type: N/A
Events:
Introduced- None
Fixed -None
Affected versions
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High