CVE-2021-33886
Dashboard / Vulnerabilities / CVE-2021-33886
CVE-2021-33886
Published: 25 Aug 2021Last Modified: 10 Mar 2026
Summary:
Details: An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
References: https://www.bbraunusa.com/en.htm, , https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Affected packages
Package
Name:
Purl:
Affected ranges
Type: N/A
Events:
Introduced- None
Fixed -None
Affected versions
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High