CVE-2019-25719
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Published:Jun 2, 2026
Last Modified:Jun 3, 2026
EPS:Jun 2, 2026
EPSS Score:0.00017
CVSS Score:8.6
Affected Products
Vendor
Product
Action
Vendor
Draeger
Product
Infinity Acute Care System
Draeger
Infinity Acute Care System
Vendor
Draeger
Product
Standalone Infinity M540 Patient Monitor
Draeger
Standalone Infinity M540 Patient Monitor
Exploits
No exploit reference
Common Weakness Enumeration
Common Attack Pattern Enumeration and Classification (CAPEC)
No CAPEC recorded yet
Common Vulnerability Scoring System
Attack Vector
Network
Adjacent
Local
Physical
Privileges Required
None
Low
High
User Interaction
None
Required
Scope
Unchanged
Changed
Confidentiality
None
Low
High
Integrity
None
Low
High
Availability
None
Low
High