CVE Feed

    Dashboard / CVE / CVE-2022-4992

    CVE-2022-4992

    Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

    Published:Jun 2, 2026
    Last Modified:Jun 3, 2026
    EPS:Jun 2, 2026
    EPSS Score:0.00045
    CVSS Score:8.6

    Affected Products

    Vendor
    Draeger
    Product
    Infinity Acute Care System
    Vendor
    Draeger
    Product
    Standalone Infinity M540 Patient Monitor

    Exploits

    No exploit reference

    Common Vulnerability Scoring System

    Attack Vector
    Network
    Adjacent
    Local
    Physical
    Privileges Required
    None
    Low
    High
    User Interaction
    None
    Required
    Scope
    Unchanged
    Changed
    Confidentiality
    None
    Low
    High
    Integrity
    None
    Low
    High
    Availability
    None
    Low
    High